Test and characterization of 400 Hamamatsu R5912-MOD photomultiplier tubes for the ICARUS T600 detector
2018
ICARUST600 will be operated as far detector of the Short Baseline Neutrino program at FNAL (USA), which foresees three liquid
argon
time projection chambersalong the Booster Neutrino Beam line to search for LSND-like
sterile neutrinosignal. The detector employs 360
photomultipliertubes, Hamamatsu model R5912-MOD, suitable for
cryogenicapplications. A total of 400 PMT were procured from Hamamatsu and tested at room temperature to evaluate the performance of the devices and their compliance to detect the liquid
argonscintillation light in the T600 detector. Furthermore 60 units were also characterized at
cryogenictemperature, in liquid
argonbath, to evaluate any parameter variation which could affect the scintillation light detection. All the tested PMTs were found to comply with the requirements of
ICARUST600 and a subset of 360 specimens were selected for the final installation in the detector.