Real-time nuclear activation detectors for measuring neutron angular distributions at the National Ignition Facility (invited)

2021
The Real Time Nuclear Activation Detector (RTNAD) array at NIF measures the distribution of 14 MeV neutrons emitted by deuterium-tritium (DT) fueled inertial confinement fusion implosions. The uniformity of the neutron distribution is an important indication of implosion symmetry and DT shell integrity. The array consists of 48 LaBr3(Ce) crystal gamma-ray spectrometers mounted outside the NIF target chamber, which continuously monitor the slow decay of the 909 keV gamma-ray line from activated 89Zr located in Zr cups surrounding each crystal. The measured decay rate dramatically increases during a DT implosion in proportion to the number of 14 MeV neutrons striking each Zr cup. The neutrons produce activated 89Zr through an (n, 2n) reaction on 90Zr, which is insensitive to low energy neutrons. The neutron flux along the detector line-of-sight at shot time is determined by extrapolating the fitted 909 keV decay curve back to shot time. Automatic analysis algorithms were developed to handle the non-stop data stream. The large number of detectors and the high statistical accuracy of the array enable the spherical harmonic modes of the neutron angular distribution to be measured up to L ≤ 4 to provide a better understanding of implosion dynamics. In addition, these data combined with measurements of the down-scattered neutrons can be used to derive fuel areal density distributions. This paper will describe the RTNAD hardware and analysis procedures.
    • Correction
    • Source
    • Cite
    • Save
    10
    References
    3
    Citations
    NaN
    KQI
    []
    Baidu
    map