A case of inappropriate implantable cardioverter defibrillator therapy induced by T-wave oversensing due to hyperkalemia

2015 
There have been reports of hyperkalemia-induced T-wave oversensing in patients with implantable cardioverter defibrillators (ICDs). However, a comparison of T-wave amplitudes and morphologies between the surface 12-lead electrocardiogram (ECG) and ICD electrogram has not been reported. We present the case of a 70-year-old man who received inappropriate ICD shocks due to hyperkalemia-induced T-wave oversensing. The T-wave amplitudes on both the ICD electrogram and 12-lead ECG corresponded and normalized after normalization of the potassium level.