A fluorogenic heterogeneous immunoassay for cardiac muscle troponin cTnI on a digital microfluidic device

2014 
We describe a fluorogenic two-site noncompetitive heterogeneous immunoassay with magnetic beads on a low-voltage digital microfluidic platform using closed electrowetting-on-dielectric (EWOD). All the steps of an enzyme-linked immunosorbent assay (ELISA) were performed on the device using 9H-(1, 3-dichloro-9, 9-dimethylacridin-2-one-7-yl) phosphate as the fluorogenic substrate for the enzyme alkaline phosphatase. The performance of the system was demonstrated with cardiac marker Troponin I (cTnI) as a model analyte in phosphate-buffered saline samples. cTnI was detected within the diagnostically relevant range with a limit of detection of 2.0 ng/mL (CV = 6.47 %). Washing of magnetic beads was achieved by movement through a narrow region of buffer bridging one drop to another with minimal fluid transfer. More than 90 % of the unbound reagents were removed after five washes. Further experiments testing human blood serum on the same platform demonstrated a sample-to-answer time at ∼18.5 min detecting 6.79 ng/mL cTnI.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    31
    References
    18
    Citations
    NaN
    KQI
    []
    Baidu
    map