Internal Calibration of the PandaX-II Detector with Radon Gaseous Sources

2020
We have developed a low-energy electron recoil (ER) calibration method with $^{220}$Rn for the PandaX-II detector. $^{220}$Rn, emanated from natural thorium compounds, was fed into the detector through the xenon purification system. From 2017 to 2019, we performed three dedicated calibration campaigns with different radon sources. The detector response to $\alpha$, $\beta$, and $\gamma$ particles was studied, with focus on low energy ER events. The amount of radioactive background, including $^{222}$Rn and thorium particulate, was measured and found to be manageable in two sources. The background contribution from $^{214}$Pb was also measured for the first time in PandaX-II with help from a $^{222}$Rn injection.