Echo-planar imaging is superior to fast spin-echo diffusion-weighted imaging for cranioplasty using titanium mesh in brain magnetic resonance imaging: a phantom study.

2021
This study aimed to compare the radiofrequency (RF) shielding effects of titanium mesh of echo-planar imaging (EPI) versus fast spin-echo (FSE) diffusion-weighted imaging (DWI) to establish a suitable sequence for patients who undergo cranioplasty and for whom titanium mesh was used in brain magnetic resonance imaging (MRI). A 1.5-T MRI scanner with clinical setting sequences was used. A phantom for the examination constructed using a sucrose solution in a plastic container was used to compare the signal attenuation (SA) ratio, area of RF shielding effect (Area), normalized absolute average deviation (NAAD), and apparent diffusion coefficient (ADC) between EPI and FSE-DWI. EPI provided significantly better SA ratio, Area, and NAAD (P < 0.01). When the number of slices increased, the RF shielding became more negative. There was no significant difference in the ADC. Regardless of the k-trajectory, EPI-DWI had a lower RF shielding effect than FSE-DWI in patients undergoing cranioplasty.