Accuracy Evaluation of Five Blood Glucose Monitoring Systems Obtained from the Pharmacy: A European Multicenter Study with 453 Subjects

2012
Abstract Background: This multicenter study was conducted to evaluate the performance of five recently introduced blood glucose (BG) monitoring (BGM) devices under daily routine conditions in comparison with the YSI (Yellow Springs, OH) 2300 Stat Plus glucose analyzer. Methods: Five hundred one diabetes patients with experience in self-monitoringof BG were randomized to use three of five different BGM devices (FreeStyle Lite® [Abbott Diabetes Care Inc., Alameda, CA], FreeStyle Freedom Lite [Abbott Diabetes Care], OneTouch® UltraEasy® [LifeScan Inc., Milpitas, CA], Accu-Chek® Aviva [ Roche Diagnostics, Mannheim, Germany], and Contour® [Bayer Vital GmbH, Leverkusen, Germany]) in a daily routine setting. All devices and strips were purchased from local regular distribution sources (pharmacies, four strip lots per device). The patients performed the finger prick and the glucose measurement on their own. In parallel, a healthcare professional performed the glucose assessment with the reference method (YSI 2300...