Qualitative Elemental Analysis by Energy Dispersive X-Ray Spectrometry

2018 
Before attempting automatic or manual peak identification, it is critical that the EDS system be properly calibrated to ensure that accurate energy values are measured for the characteristic X-ray peaks. Follow the vendor’s recommended procedure to rigorously establish the calibration. The calibration procedure typically involves measuring a known material such as copper that provides characteristic X-ray peaks at low photon energy (e.g., Cu L3-M5 at 0.928 keV) and at high photon energy (Cu K-L3 at 8.040 keV). Alternatively, a composite aluminum-copper target (e.g., a copper penny partially wrapped in aluminum foil and continuously scanned so as to excite both Al and Cu) can be used to provide the Al K-L3 (1.487 keV) as the low energy peak and Cu K-L3 for the high energy peak. After calibration, peaks occurring within this energy range (e.g., Ti K-L3 at 4.508 keV and Fe K-L3 at 6.400 keV) should be measured to confirm linearity. A well-calibrated EDS should produce measured photon energies within ±2.5 eV of the ideal value. Low photon energy peaks below 1 keV photon energy should also be measured, for example, O K (e.g., from MgO) and C K. For some EDS systems, non-linearity may be encountered in the low photon energy range. Figure 18.1 shows an EDS spectrum for CaCO3 in which the O K peak at 0.523 keV is found at the correct energy, but the C K peak at 0.282 keV shows a significant deviation below the correct energy due to non-linear response in this range.