Diagnostic accuracy of Panbio™ rapid antigen tests on oropharyngeal swabs for detection of SARS-CoV-2

2021
BackgroundAntigen-detecting rapid diagnostic tests (Ag-RDTs) for the detection of SARS-CoV-2 offer new opportunities for testing in the context of the COVID-19 pandemic. Nasopharyngeal swabs (NPS) are the reference sample type, but oropharyngeal swabs (OPS) may be a more acceptable sample type in some patients. MethodsWe conducted a prospective study in a single screening center to assess the diagnostic performance of the Panbio COVID-19 Ag Rapid Test (Abbott) on OPS compared with reverse-transcription quantitative PCR (RT-qPCR) using NPS. Results402 outpatients were enrolled in a COVID-19 screening center, of whom 168 (41.8%) had a positive RT-qPCR test. The oropharyngeal Ag-RDT sensitivity compared to nasopharyngeal RT-qPCR was 81% (95%CI: 74.2-86.6). Two false positives were noted out of the 234 RT-qPCR negative individuals, which resulted in a specificity of 99.1% (95%CI: 96.9-99.9) for the Ag-RDT. For cycle threshold values [≤] 26.7 ([≥] 1E6 SARS-CoV-2 genomes copies/mL, a presumed cut-off for infectious virus), 96.3% sensitivity (95%CI: 90.7-99.0%) was obtained with the Ag-RDT using OPS. InterpretationBased on our findings, the diagnostic performance of the Panbio Covid-19 RDT with OPS samples meet the criteria required by the WHO for Ag-RDTs (sensitivity[≥]80% and specificity [≥]97%).