Quantification of Isomer-Resolved Iodide CIMS Sensitivity and Uncertainty Using a Voltage Scanning Approach

2021
Abstract. Chemical ionization mass spectrometry (CIMS) using iodide as a reagent ion has been widely used to classify organic compounds in the atmosphere by their elemental formula. Unfortunately, calibration of these instruments is challenging due to a lack of commercially available standards for many compounds, which has led to the development of methods for estimating CIMS sensitivity. By coupling a Thermal desorption Aerosol Gas chromatograph (TAG) simultaneously to a flame ionization detector (FID) and an iodide CIMS, we use the individual particle-phase analytes, quantified by the FID, to examine the sensitivity of the CIMS and its variability between isomers of the same elemental formula. Iodide CIMS sensitivities of isomers within a formula are found to generally vary by one order of magnitude with a maximum deviation of two orders of magnitude. Furthermore, we compare directly measured sensitivity to a method of estimating sensitivity based on declustering voltage (i.e., “voltage scanning”). This approach is found to carry high uncertainties for individual analytes (half to one order of magnitude), but represents a central tendency that can be used to estimate the sum of analytes with reasonable error (~30 % differences between predicted and measured moles). Finally, GC retention time, which is associated with vapor pressure and chemical functionality of an analyte, is found to qualitatively correlate with iodide CIMS sensitivity, but the relationship is not close enough to be quantitatively useful and could be explored further in the future as a potential calibration approach.
    • Correction
    • Source
    • Cite
    • Save
    23
    References
    0
    Citations
    NaN
    KQI
    []
    Baidu
    map