How to Validate in silico Deployment of Coronary Stents: Strategies and Limitations in the Choice of Comparator

2021
This work aims at proposing and discussing useful indications to all those who need to validate a numerical model of coronary stent deployment. The proof of the reliability of a numerical model is becoming of paramount importance in the era of in silico trials. Recently the ASME V&V Standard Committee for medical devices prepared the V&V 40 standard document that provides a framework that guides users in establishing and assessing the relevance and adequacy of verification and validation activities performed for proving models' credibility. To the authors' knowledge, only a few examples of the application of the V&V 40 framework are available in the literature applied to medical devices, but none about stents. Specifically, in this study, the authors wish to emphasize the choice of a relevant set of experimental activities to provide data for the validation of computational models aiming to predict coronary stent deployment. Attention is focused on the use of ad hoc 3D-printed mock vessels in the validation plan, which could allow evaluating aspects of clinical relevance in a representative but controlled environment.
    • Correction
    • Source
    • Cite
    • Save
    21
    References
    1
    Citations
    NaN
    KQI
    []
    Baidu
    map