Clinical Implementation of Volumetric Intensity-Modulated Arc Therapy (VMAT) with ERGO++

2010
Background and Purpose:Volumetric modulated arc therapy (VMAT) has the potential to deliver dose distributions comparable to the established intensity-modulated radiotherapy techniques for a multitude of target paradigms. Prior to implementing VMAT into their clinical routine in December 2008, the authors evaluated the dose calculation/delivery accuracy of 24 sample VMAT plans (prostate and anal cancertarget paradigms) with film and ionization dosimetry. After the start of the clinical program, in vivo measurements with a rectal probe were performed.Material and Methods:The VMAT plans were generated by the treatment-planning system (TPS) ERGO++ (Elekta, Crawley, UK) and transferred to a phantom. Film dosimetrywas performed with Kodak EDR2 films, and evaluated with dose profilesand γ-index analysis. Appropriate ionization chambers were used for absolute dose measurements in the phantom and for in vivo measurements. The ionization chamber was used with localization of the measurement volume based on positioning cone-beam computed tomography.Results:Plans were transferred from ERGO++ to the record and verify (R&V) system/linear accelerator (linac). The absolute dose deviations recorded with the ionization chamber were 1.74% ± 1.62% across both indications. The γ-index analysis of the film dosimetryshowed no deviation > 3%/3 mm in the high-dose region. On in vivo measurements, a deviation between calculation and measurement of 2.09% ± 2.4% was recorded, when the chamber was successfully positioned in the high-dose region.Conclusion:VMAT plans can be planned and treated reproducibly in high quality after the commissioning of the complete delivery chain consisting of TPS, R&V system and linac. The results of the individual plan verification meet the commonly accepted requirements. The first in vivo measurements confirm the reproducible precision of the delivered dose during clinical treatments.
    • Correction
    • Source
    • Cite
    • Save
    54
    References
    40
    Citations
    NaN
    KQI
    []
    Baidu
    map