Low dose cone beam CT for paediatric image-guided radiotherapy: Image quality and practical recommendations

2021 
Abstract Purpose Cone beam CT (CBCT) is used in paediatric image-guided radiotherapy (IGRT) for patient setup and internal anatomy assessment. Adult CBCT protocols lead to excessive doses in children, increasing the risk of radiation-induced malignancies. Reducing imaging dose increases quantum noise, degrading image quality. Patient CBCTs also include ‘anatomical noise’ (e.g. motion artefacts), further degrading quality. We determine noise contributions in paediatric CBCT, recommending practical imaging protocols and thresholds above which increasing dose yields no improvement in image quality. Methods and materials Sixty CBCTs including the thorax or abdomen/pelvis from 7 paediatric patients (aged 6–13 years) were acquired at a range of doses and used to simulate lower dose scans, totalling 192 scans (0.5–12.8 mGy). Noise measured in corresponding regions of each patient and a 10-year-old phantom were compared, modelling total (including anatomical) noise, and quantum noise contributions as a function of dose. Contrast-to-noise ratio (CNR) was measured between fat/muscle. Soft tissue registration was performed on the kidneys, comparing accuracy to the highest dose scans. Results Quantum noise contributed  1 mGy. Conclusion Anatomical noise dominates quantum noise in paediatric CBCT. Appropriate soft tissue contrast and registration accuracy can be achieved for doses as low as 1 mGy. Increasing dose above 1 mGy holds no benefit in improving image quality or registration accuracy due to the presence of anatomical noise.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    50
    References
    0
    Citations
    NaN
    KQI
    []
    Baidu
    map