Repeatability of transient elastography in children.

2020 
Poorly performing diagnostic tests can impact patient safety. Clinical investigations must have good precision and diagnostic accuracy before widespread use in clinical practice. Transient elastography (TE) measures liver stiffness, a surrogate marker of liver fibrosis in adults and children. Studies to evaluate its repeatability and reproducibility (precision) in children are limited. Our aim was to determine (i) the normal range of TE measurements and (ii) the repeatability and reproducibility of TE in healthy children. TE was performed in 257 healthy children, of whom 235 (91%, mean age 11.7 years, standard deviation (SD) 2.51, 107 were males (45.5%)) had two valid TE measurements performed, at least 24 h apart, by two operators under similar circumstances. High-quality TE images were obtained for each examination. The normal range of TE was 2.88–6.52 kPa. The mean difference between paired measurements was 0.044 (SD 0.4). The 95% limits of agreement ranged from −0.8 to +0.76 kPa for repeat measurements. There was a difference of >1 kPa between measurements in 61/235 (25.9%) children. The lack of precision was similar across all age groups. This study demonstrates that TE does not have acceptable precision in healthy children, because random measurement variation results in the lack of agreement between paired measurements.