Comparative analysis of radiation dose and image quality between organ dose modulation and 3D smart mA modulation during head-neck CT angiography

2019
OBJECTIVE: To assess the difference in absorbed organ dose and image quality for head-neck CT angiography using organ dose modulation compared with 3D smart mA modulation in different body mass indices (BMIs) using an adaptive statistical iterative reconstruction (ASiR-V) algorithm. METHODS: Three hundred patients underwent head-neck CTA were equally divided into three groups: A (18.5 kg/m2≦BMI   0.05). However, there is significant difference of ED values (P  <  0.05) . For example, in subgroup A1 mean ED values are 15.30% and 23.66% lower than those in subgroup A2 at thyroid gland and eye lens, respectively. Similar patterns also exist in groups B (B1 vs. B2) and C (C1 vs. C2). CONCLUSIONS: Using organ dose modulation and applying the ASiR-V algorithm can more effectively reduce the radiation dose in head-neck CT angiography than using 3D smart mA modulation, while maintaining image quality. Thus, using organ-based dose modulation has the additional benefit of reducing dose to the thyroid gland and eye lens.