Evaluation of the Precision Xtra meter for monitoring blood β-hydroxybutyrate concentrations in late-gestation ewes:

2019 
Blood samples were collected from late-gestation ewes to determine the agreement of a point-of-care (POC) Precision Xtra meter and a standard laboratory test for β-hydroxybutyrate (BHBA). Fresh whole blood samples were immediately tested with the POC instrument, and serum samples were analyzed with a standard commercial biochemical analyzer. Ewes were classified as having ketonemia if their BHBA concentrations were ≥800 µmol/L. Scatter plots, paired t-tests, Bland–Altman limits of agreement, and Gwet AC1 tests were used to compare results. The 2 tests had very good agreement. The values between instruments were not statistically different based on paired t-tests (p = 0.312). The intercept and slope of a linear mixed model, containing the standard test results as an outcome and the POC meter results as a predictor, were 0.02 (95% CI: 0.00, 0.04) and 0.98 (95% CI: 0.96, 1.01), respectively. When the samples were classified into ketonemic classes (non-ketonemic and ketonemic) based on BHBA concentrations obt...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    21
    References
    2
    Citations
    NaN
    KQI
    []
    Baidu
    map