Cement-Augmented Carbon Fiber-Reinforced Pedicle Screw Instrumentation for Spinal Metastases: Safety and Efficacy.

2021 
Objective To investigate the complication rates and long-term implant failure rates in a monocentric study of a consecutive cohort of patients with thoracolumbar spinal metastases after posterior instrumentation with a fenestrated carbon fiber–reinforced poly-ether-ether-ketone (CFRP) pedicle screw system. Methods We retrospectively reviewed demographics, Karnofsky Performance Status Scale scores, complications, and implant failure rates. Results Between June 2016 and November 2019, 51 consecutive patients underwent cement-augmented CFRP pedicle screw instrumentation at our institution. Mean age was 68 years (standard deviation 10.5), the median preoperative Karnofsky Performance Status Scale of 80 increased to 90 postoperatively (P = 0.471). Most common primary entities were breast (25.5%), lung (15.7%), and prostate (13.7%) cancers. Of 428 placed screws, 293 (68.5%) were augmented with polymethylmethacrylate, a mean 6 per patient (standard deviation ±2). Screws were inserted via a minimally invasive system technique in 54.9% of cases. In total, 11.8% of patients had immediate postoperative sequelae related to the cement. Pulmonary cement embolisms were noted in 3 patients, 2 had paravertebral extravasation, and 1 had an embolism into a segmental artery. Of these 6, 2 patients with pulmonary embolisms reported related symptoms. Follow-up was available for 80.4%. After a mean 9.8 months, screw loosening was noted in 11.8% of cases on computed tomography, although it was asymptomatic in all but 1 patient. Screw pull-out did not occur. Neither cement-related (P = 0.353) nor general complication rates (P = 0.507) differed significantly between open and minimally invasive system techniques. Conclusions Percutaneous cement-augmented CFRP pedicle screw instrumentation facilitates artifact-reduced postoperative imaging, while maintaining a risk profile and implant failure rates comparable to conventional metallic instrumentation.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    46
    References
    0
    Citations
    NaN
    KQI
    []
    Baidu
    map