High-voltage micro discharge in ultra-low background /sup 3/He proportional counters

1999
This paper discusses the phenomenon of surface micro discharge induced by high voltageand techniques for the reduction and discrimination of such breakdowns in ultra-low background proportional counters. An array of ultra-low background /sup 3/He-filled proportional counterswill measure the neutral-currentinteraction rate of all active neutrino species in the Sudbury Neutrino Observatory. The sensitivity of these Neutral CurrentDetectors and their stringent background criteria make it essential to minimize all spurious pulses including signals induced by high voltage. Such pulses can originate from microscopic surface discharges in the various proportional countercomponents. Studies have shown that this discharge effect occurs mainly at interfaces and in microscopic voids between dielectric surfaces, on contaminated surfaces, and on surfaces with imperfections. Because of its occurrence in various materials and environments, micro discharge is a concern for all low-background detectors that operate under high voltage.