SELF-REPORTED SYSTEMIC COMPLAINTS IN PATIENTS WITH METAL-ON-METAL HIP ARTHROPLASTY
2018
BackgroundThe second generation metal-on-metal (MoM) prosthesis of the hip became a worldwide success in the 90s. However, after the placement of a MoM prosthesis the cobalt ion concentrations raise significantly. This may lead to systemic complaints and even cobalt toxicity.MethodsSixty-one patients (26F/35M) with both an unilateral and bilateral resurfacing or large-head MoM (LHMoM) hip prosthesis were included. At last follow-up (5.77 ± 1.57 yrs) cobalt concentrations in the blood were determined by ICP-MS. Based on the known cobalt toxicity symptoms we developed a non-validated questionnaire. Analysis was done on two groups; a low cobalt concentration group and a high cobalt concentration group. We used 170 nmol/L as the upper limit of well functioning prosthesis as defined by the Dutch Orthopaedic Society (NOV). Independent samples T test and Pearson correlation coefficient were done.ResultsIn 61 patients, 70 prosthesis were implanted: 44 resurfacings and 26 LHMoM. Mean cobalt and chromium ion concen...