Body surface and body core temperatures and their associations to haemodynamics: The BOSTON-I-study: Validation of a thermodilution catheter (PiCCO) to measure body core temperature and comparison of body surface temperatures to thermodilutionderived Cardiac Index

2020 
Assessment of peripheral perfusion and comparison of surface and body core temperature (BST; BCT) are diagnostic cornerstones of critical care. Infrared non-contact thermometers facilitate the accurate measurement of BST. Additionally, a corrected measurement of BST on the forehead provides an estimate of BCT (BCT_Forehead). In clinical routine BCT is measured by ear thermometers (BCT_Ear). The PiCCO-device (PiCCO: Pulse contour analysis) provides thermodilution-derived Cardiac Index (CI_TD) using an arterial catheter with a thermistor tip in the distal aorta. Therefore, the PiCCO-catheter might be used for continuous BCT-measurement (BCT_PiCCO) in addition to intermittent CI-measurement. To the best of our knowledge, BCT_PiCCO has not been validated compared to standard techniques of BCT-measurement including measurement of urinary bladder temperature (BCT_Bladder). Therefore, we compared BCT_PiCCO to BCT_Ear and BCT_Bladder in 52 patients equipped with the PiCCO-device (Pulsion; Germany). Furthermore, this setting allowed to compare different BSTs and their differences to BCT with CI_TD. BCT_PiCCO, BCT_Ear (ThermoScan; Braun), BCT_Bladder (UROSID; ASID BONZ), BCT_Forehead and BSTs (Thermofocus; Tecnimed) were measured four times within 24h. BSTs were determined on the great toe, finger pad and forearm. Immediately afterwards TPTD was performed to obtain CI_TD. 32 (62%) male, 20 (38%) female patients; APACHE-II 23.8 ±8.3. Bland-Altman-analysis demonstrated low bias and percentage error (PE) values for the comparisons of BCT_PiCCO vs. BCT_Bladder (bias 0.05 ±0.27° Celsius; PE = 1.4%), BCT_PiCCO vs. BCT_Ear (bias 0.08 ±0.38° Celsius; PE = 2.0%) and BCT_Ear vs. BCT_Bladder (bias 0.04 ±0.42° Celsius; PE = 2.2). While BCT_PiCCO, BCT_Ear and BCT_Bladder can be considered interchangeable, Bland-Altman-analyses of BCT_Forehead vs. BCT_PiCCO (bias =-0.63 ±0.75° Celsius; PE = 3.9%) Celsisus, BCT_Ear (bias = -0.58 ±0.68° Celsius; PE = 3.6%) and BCT_Bladder (bias = -0.55 ±0.74° Celsius; PE = 3.9%) demonstrate a substantial underestimation of BCT by BCT_Forehead. BSTs and differences between BCT and BST (DCST) significantly correlated with CI_TD with r-values between 0.230 and 0.307 and p-values between 0.002 and p < 0.001. The strongest association with CI_TD was found for BST_forearm (r = 0.307; p < 0.001). In a multivariate analysis regarding CI_TD and including biometric data, BSTs and and their differences to core-temperatures (DCST), only higher temperatures on the forearm and the great toe, young age, low height and male gender were independently associated with CI_TD. The estimate of CI based on this model (CI_estimated) correlated with CI_TD (r = 0.594; p < 0.001). CI_estimated provided large ROC-areas under the curve (AUC) regarding the critical thresholds of CI_TD ≤ 2.5 L/min/m2 (AUC = 0.862) and CI_TD ≥ 5.0 L/min/m2 (AUC = 0.782). 1.) BCT_PiCCO, BCT_Ear and BCT_Bladder are interchangeable. 2.) BCT_Forehead significantly underestimates BCT by about 0.5° Celsius. 3.) All measured BSTs and DCSTs were significantly associated with CI_TD. 4.) CI_estimated is promising, in particular for the prediction of critical thresholds of CI.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    0
    Citations
    NaN
    KQI
    []
    Baidu
    map