SU‐F‐T‐261: Reconstruction of Initial Photon Fluence Based On EPID Images
2016
Purpose: Verifying an algorithm to reconstruct relative initial photon
fluencefor clinical use. Clinical EPID and CT images were acquired to reconstruct an external photon radiation treatment field. The reconstructed initial photon
fluencecould be used to verify the treatment or calculate the applied dose to the patient. Methods: The acquired EPID images were corrected for scatter caused by the patient and the EPID with an
iterative reconstructionalgorithm. The transmitted photon
fluencebehind the patient was calculated subsequently. Based on the transmitted
fluencethe initial photon
fluencewas calculated using a back-projection algorithm which takes the patient geometry and its energy dependent linear attenuation into account. This attenuation was gained from the acquired cone-beam CT or the planning CT by calculating a water-equivalent radiological thickness for each irradiation direction. To verify the algorithm an inhomogeneous phantom consisting of three inhomogeneities was irradiated by a static 6 MV photon field and compared to a reference flood field image. Results: The mean deviation between the reconstructed relative photon
fluencefor the inhomogeneous phantom and the flood field EPID image was 3% rising up to 7% for off-axis
fluence. This was probably caused by the used clinical EPID calibration, which flattens the inhomogeneous
fluenceprofile of the beam. Conclusion: In this clinical experiment the algorithm achieved good results in the center of the field while it showed high deviation of the lateral
fluence. This could be reduced by optimizing the EPID calibration, considering the off-axis differential energy response. In further progress this and other aspects of the EPID, eg. field size dependency, CT and dose calibration have to be studied to realize a clinical acceptable accuracy of 2%.
Keywords:
-
Correction
-
Source
-
Cite
-
Save
0
References
0
Citations
NaN
KQI