Log file-based patient dose calculations of double-arc VMAT for head-and-neck radiotherapy
2018
Abstract Purpose The
logfile-based method cannot display dosimetric changes due to linac component miscalibration because of the insensitivity of
logfiles to linac component miscalibration. The purpose of this study was to supply dosimetric changes in
logfile-based patient dose calculations for double-arc volumetric-modulated arc therapy (VMAT) in head-and-neck
cases.
Methodsand materials Fifteen head-and-neck cases participated in this study. For each case, treatment planning system (TPS) doses were produced by double-arc and single-arc VMAT. Miscalibration-simulated
logfiles were generated by inducing a leaf miscalibration of ±0.5 mm into the
logfiles that were acquired during VMAT irradiation. Subsequently, patient doses were estimated using the miscalibration-simulated
logfiles. Results For double-arc VMAT, regarding planning target volume (PTV), the change from TPS dose to miscalibration-simulated
logfile dose in D mean was 0.9 Gy and that for tumor control probability was 1.4%. As for organ-at-risks (OARs), the change in D mean was mean and
radiobiologicalmetrics ( P Conclusions Using the
logfile-based method for PTV and OARs, the
logfile-based method estimate of patient dose using the double-arc VMAT has accuracy comparable to that obtained using the single-arc VMAT.
Keywords:
-
Correction
-
Source
-
Cite
-
Save
26
References
1
Citations
NaN
KQI