Mass test of AdvanSiD model ASD-NUV3S-P SiliconPMs for the Pixel Timing Counter of the MEG II experiment

2017 
The MEG II Timing Counter will measure the positron time of arrival with a resolution of 30 ps relying on two arrays of scintillator pixels read out by 6144 Silicon Photomultipliers (SiPMs) from AdvanSiD. They must be characterized, measuring their breakdown voltage, to assure that the gains of the SiPMs of each pixel are as uniform as possible, to maximize the pixel resolution. To do this an automatic test system that can measure sequentially the parameters of 32 devices has been developed.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    12
    References
    2
    Citations
    NaN
    KQI
    []
    Baidu
    map