Thermal crosstalk of X-ray transition-edge sensor micro-calorimeters under frequency domain multiplexing readout

2021
We have measured and characterized the thermal crosstalk in two different arrays of transition-edge sensors (TES) X-ray micro-calorimeters with frequency-domain multiplexing (FDM) readout. The TES arrays are fabricated at SRON and are a 88 and a 3232 "kilo-pixel" uniform array. The amount of crosstalk is evaluated as the ratio between the averaged crosstalk signal and the X-ray pulse amplitudes. The crosstalk ratios for our detectors are compliant with the requirements for future X-ray space missions, such as Athena X-IFU, with a nearest-neighbour thermal crosstalk ratio at a level of 10-4, with a highest crosstalk ratio of 410-4 for the kilo-pixel array (worst case, center of array) and 110-4 for the 88 array, with a margin of improvement achievable by optimizing the Cu metallization and the width of the Si supporting structures (muntins) in the backside of the TES array chip. Based on the measured crosstalk ratios, we have estimated the impact on the spectral resolution by means of noise equivalent power (NEP) considerations and a Monte Carlo simulation, finding an average degradation in quadrature of less than 40 meV, compliant with the < 0.2 eV requirement for Athena X-IFU.