Mapping and Isolation of the Pulmonary Veins Using the PVAC Catheter
2010
Objectives:We aimed to investigate the feasibility, efficacy, and safety of the pulmonary vein ablation catheter (PVAC) catheter (a novel multielectrode catheter using duty-cycled bipolar and unipolar radiofrequency energy, Medtronic, Minneapolis, MN, USA) to completely isolate the pulmonary veins (PVs).
Methods:Twenty-seven patients (60 ± 8 years) with paroxysmal atrial fibrillation (AF) underwent PV isolation with the PVAC catheter. PVAC was used for both mapping and isolation of the PVs (PVAC-guided ablation). After PVAC ablation, presence/absence of PV potentials (PVP) was verified using a conventional circular mapping catheter. In case of residual PVP on the circular catheter, PVAC ablation was continued.
Results:After PVAC-guided ablation 99 of 106 PVs (93%) and 21 of 27 patients (78%) were proven to be isolated. Failure to isolate was due to a mapping failure in four right-sided PVs and a true ablation failure in three right-sided PVs. After continued PVAC ablation, 103 of 106 PVs (97%) and 25 of 27 patients (93%) were shown to be isolated. The total procedural time from femoral vein access to complete catheter withdrawal was 176 ± 25 minutes. The actual dwelling-time of the PVAC within the left atrium was 102 ± 37 minutes. Esophageal T° rise to >38.5° occurred in nine of 19 monitored patients (47%).
Conclusions:(1) PVAC-guided ablation (i.e., mapping and ablation with a single catheter) results in isolation of all PVs in 73% of the patients. (2) An additional circular mapping catheter is required to increase complete isolation rate to 93% of the patients. (3) Given the esophageal T° rise in almost 50% of patients, safety precautions are needed. (PACE 2010; 33:168–178)
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
11
References
38
Citations
NaN
KQI