On the linearity of the MINOS light-injection calibration system
2004
The
MINOSlight-injection calibration system has been fully documented in a previous article (Nucl. Instr. and Meth. A 492 (2002) 353). Upon commissioning in the
MINOSdetectors, the system was found to give a non-linear response to variations in the intensity of injected light. The source of this non-linearity has been traced to a small change in the spectrum of the injected light as a function of the current applied to the original blue LEDs, in combination with a rapidly varying spectral response function of the wavelength-shifting fibre used in the detector. Both aspects of the problem have been addressed successfully by use of LEDs with different spectral characteristics, and the system now has a linear response.
Keywords:
-
Correction
-
Source
-
Cite
-
Save
3
References
9
Citations
NaN
KQI